Al, As, B, Ba, Ca, Cr, Cu, Fe, Ga, Ge, In, K, Mg, Mn, Na, P, Sb, Ta, Ti, Tl
-
Al, B, Ba, Ca, Cr, Cu, Fe, Ga, Ge, In, K, Li, Mg, Mn, Na, Ni, P, Sn, Ta, Ti
GNM-M201137-2013 | 100ml
-
Al, B, Ba, Cr, Cu, Fe, Ga, Ge, In, K, Li, Mg, Mn, Na, Ni, P, Sn, Ta, Ti
GNM-M191136-2013 | 100ml
-
Al, B, Ba, Be, Ca, Cr, Cu, Fe, K, Li, Mg, Mn, Na, Nb, P, Ta, Ti
GNM-M170218-2013 | 100ml
-
Al, B, Ba, Ca, Cr, Cu, Fe, Ge, K, Li, Mg, Mn, Na, Ni, P, Ti, Zr
GNM-M1711175-2013 | 100ml
-
Ag, Al, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Ga,Ge,K, Li, Mg, Mn, Na, Ni, Pb, Sb, Sn, Ta, Ti, V, Zn
GNM-M256797-2013 | 100ml
-
Al, B, Ba, Ca, Cr, Cu, Fe, K, Li, Mg, Mn, Na, P, Ti
GNM-M140164-2013 | 100ml
-
Ag Al As B Ba Bi Ca Cd Co Cr Cu Fe Ga Ge In K Li Mg Mn Mo Na Ni Pb Sn Sr Ta Ti Tl W Zn
GNM-M3010766-2013 | 100ml
介质及浓度 : 10%HNO3, 痕量HF, 有效期(月) : 12;
您正在浏览的产品:20种金属元素混标溶液/铝(Al)砷(As)硼(B)钡(Ba)钙(Ca)铬(Cr)铜(Cu)铁(Fe)镓(Ga)锗(Ge)铟(In)钾(K)镁(Mg)锰(Mn)钠(Na)磷(P)锑(Sb)钽(Ta)钛(Ti)铊(Tl)/介质:10%HNO3, 痕量HF
手机版:20种金属元素混标溶液/铝(Al)砷(As)硼(B)钡(Ba)钙(Ca)铬(Cr)铜(Cu)铁(Fe)镓(Ga)锗(Ge)铟(In)钾(K)镁(Mg)锰(Mn)钠(Na)磷(P)锑(Sb)钽(Ta)钛(Ti)铊(Tl)/介质:10%HNO3, 痕量HF
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
砷(As):0.02μg/L
硼(B):0.10μg/L
其他元素检出限0.01~0.50μg/L,定量限为检出限的3倍
2. 每10个样品插入1个空白对照
3. 加标回收率控制在85%~115%
4. 内标元素回收率需≥90%
2. 加入5mL硝酸+1mL氢氟酸微波消解
3. 180℃消解30分钟后定容至50mL
4. 加入1μg/L铑内标溶液
5. ICP-MS分析前用调谐液优化仪器参数
2. 含HF介质需使用耐氢氟酸进样系统
3. 钛(Ti)含量>0.5%时需稀释后测定
4. 每4小时需重新校准标准曲线
钠(Na):0.0005%
磷(P):0.0008%
其他元素定量限为0.001%~0.003%
2. 每批次带基体加标样(回收率90%~110%)
3. 连续分析时每2小时插入质控样
4. 重复样相对偏差≤5%
2. 加入15mL硝酸+1mL氢氟酸(含硅样品)
3. 电热板150℃溶解至澄清
4. 转移定容至100mL容量瓶
5. ICP-AES分析选择推荐谱线:Al 396.15nm、B 249.77nm、Ti 334.94nm
2. 高含量元素(>5%)需选择次灵敏谱线
3. 氢氟酸介质必须使用耐HF雾化器
4. 钠、钾元素需使用塑料器皿防止污染
锑(Sb):0.03μg/L
铊(Tl):0.002μg/L
其他元素定量限0.05~2.00μg/L
2. 使用国家有证水质标准物质验证
3. 内标回收率控制在85%~115%
4. 连续校准核查(CCC)每4小时一次
2. 加入硝酸使样品酸度为1%
3. 加入在线内标混合溶液
4. 优化仪器参数:RF功率1550W,载气流量1.05L/min
5. 选择无干扰同位素:75As、121Sb、205Tl等
2. 砷需监测40Ar35Cl干扰
3. 含有机质样品需硝酸消解处理
4. 铊(Tl)等痕量元素需专用保存瓶
以上信息仅供参考,请以相应标准的原文为准!