Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Mg, Mn, Na, Ni, Pb, Se, Sr, Tl, V, Zn
-
Al As B Ba Ca Cd Co Cr Cu Fe K Mg Mn Na Ni Pb Se Sr Tl V Zn
GNM-M2111885-2013 | 100ml
-
Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Mg, Mn, Mo, Na, Ni, Pb, Sb, Se, Sn, Sr, Tl, V, Zn
GNM-M257470-2013 | 100ml
-
Al, As, Ba, Ca, Cd, Co, Cu, Fe, K, Mg, Mn, Na, Ni, Pb, Se, Sr, Tl, V, Zn
GNM-M195852-2013 | 100ml
-
Ag, Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, K, Li, Mg, Mn, Na, Ni, Pb, Se, Sr, Tl, V, Zn
GNM-M235327-2013 | 100ml
-
Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Li, Mg, Mn, Na, Ni, Pb, Se, V, Zn
GNM-M216809-2013 | 100ml
-
Al As Ba Ca Cd Co Cr Cu Fe K Mg Mn Na Ni Pb Se Sr V Zn
GNM-M1912105-2013 | 100ml
-
Ag, Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, K, Mg, Mn, Na, Ni, Pb, Se, Sr, V, Zn
GNM-M217540-2013 | 100ml
介质及浓度 : 10%HNO3, 痕量 HCl, 有效期(月) : 12;
您正在浏览的产品:22种金属元素混标溶液/铝(Al)砷(As)硼(B)钡(Ba)钙(Ca)镉(Cd)钴(Co)铬(Cr)铜(Cu)铁(Fe)汞(Hg)钾(K)镁(Mg)锰(Mn)钠(Na)镍(Ni)铅(Pb)硒(Se)锶(Sr)铊(Tl)钒(V)锌(Zn)/介质:10%HNO3, 痕量 HCl
手机版:22种金属元素混标溶液/铝(Al)砷(As)硼(B)钡(Ba)钙(Ca)镉(Cd)钴(Co)铬(Cr)铜(Cu)铁(Fe)汞(Hg)钾(K)镁(Mg)锰(Mn)钠(Na)镍(Ni)铅(Pb)硒(Se)锶(Sr)铊(Tl)钒(V)锌(Zn)/介质:10%HNO3, 痕量 HCl
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
As/Cd/Pb/Se:≤0.1 μg/L
Al/Fe/Zn:≤5 μg/L
定量限通常为检出限的3倍
2. 至少10%平行样,相对偏差≤20%
3. 加标回收率控制在80%~120%
4. 使用有证标准物质(如GBW08607水质标样)验证
2. 内标加入:上机前加入内标溶液(终浓度10~100 μg/L)
3. 仪器优化:每日使用调谐液优化质谱参数,确保氧化物产率(CeO⁺/Ce⁺)<3%
4. 干扰校正:对ArCl⁺(干扰As)、氧化物(干扰V/Cr)等进行数学校正
2. Hg元素需在样品中加入Au稳定剂(0.2% Au)防止吸附
3. 当总溶解固体>0.5%时需稀释样品或采用高基体进样系统
2. 加标浓度需覆盖待测物浓度的0.5~2倍
3. 每20个样品插入1个质控样(如GBW10010米粉标样)
4. 连续监测内标回收率(70%~130%)
2. 介质转换:消解后需赶酸至近干,用5%硝酸定容
3. 内标选择:推荐使用Li/Sc/Ge/Y/In/Tb/Lu混合内标
4. 硒/砷检测:建议使用动态反应池(DRC)消除Ar₂⁺干扰
2. 您使用的22元素混标可用于曲线绘制,但需注意B、Hg不适用此法(检出限不足)
3. 建议轴向观测提高Na/K等痕量元素灵敏度
2. 消解程序:采用梯度升温(5min→120℃, 5min→180℃, 20min→200℃)
3. 介质要求:最终定容溶液为5% HNO₃(与您使用的混标介质兼容)
以上信息仅供参考,请以相应标准的原文为准!