Al, As, B, Ba, Bi, Ca, Cd, Co, Cr, Cu, Fe, Ga, In, K, Mg, Mn, Mo, Na, Ni, Pb, Sc, Se, Te, V, Zn, Zr
-
Al, As, B, Ba, Bi, Ca, Cd, Co, Cr, Cu, Fe, Ga, In, K, Mg, Mn, Mo, Ni, Pb, Sc, Se, Te, V, Zn, Zr
GNM-M250435-2013 | 100ml
-
Al,As,B,Ba,Bi,Ca,Cd,Co,Cr,Cu,Fe,Ga,In,K,Mg,Mn,Mo,Na,Ni,Pb,Sb,Se,Sr,Te,V,Zn,Zr
GNM-M279513-2013 | 100ml
-
Al, As, B, Ba, Bi, Ca, Cd, Co, Cr, Cu, Fe, Ga, Hg, In, K, Mg, Mn, Mo, Na, Ni, Pb, Sb, Sc, Se, Sn, Sr, Te, V, Zn, Zr
GNM-M305289-2013 | 100ml
-
Al,As,B,Ba,Bi,Ca,Cd,Co,Cr,Cu,Fe,Ga,Hg,In,K,Mg,Mn,Mo,Na,Ni,Pb,Sb,Se,Sc,Sr,Te,V,Zn
GNM-M289452-2013 | 100ml
-
Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Ga, Hg, In, K Mg, Mn, Mo, Na, Ni, Pb, Sb, Se, Sr, Te, V, Zn, Zr
GNM-M276688-2013 | 100ml
-
Ag,Al,As,B,Ba,Bi,Ca,Cd,Co,Cr,Cu,Fe,Ga,K,Li,Mg,Mn,Na,Ni,Pb,Se,Sr,Te,V,Zn
GNM-M259417-2013 | 100ml
-
Al, As, B, Ba, Bi, Ca, Cd, Co, Cr, Cu, Fe, Ga, In, K, Mg, Mn, Mo, Na, Ni, Pb
GNM-M200322-2013 | 100ml
介质及浓度 : 10%HNO3,痕量 HCl,HF, 有效期(月) : 12;
您正在浏览的产品:26种金属元素混标溶液/铝(Al)砷(As)硼(B)钡(Ba)铋(Bi)钙(Ca)镉(Cd)钴(Co)铬(Cr)铜(Cu)铁(Fe)镓(Ga)铟(In)钾(K)镁(Mg)锰(Mn)钼(Mo)钠(Na)镍(Ni)铅(Pb)钪(Sc)硒(Se)碲(Te)钒(V)锌(Zn)锆(Zr)/介质:10%HNO3,痕量 HCl,HF
手机版:26种金属元素混标溶液/铝(Al)砷(As)硼(B)钡(Ba)铋(Bi)钙(Ca)镉(Cd)钴(Co)铬(Cr)铜(Cu)铁(Fe)镓(Ga)铟(In)钾(K)镁(Mg)锰(Mn)钼(Mo)钠(Na)镍(Ni)铅(Pb)钪(Sc)硒(Se)碲(Te)钒(V)锌(Zn)锆(Zr)/介质:10%HNO3,痕量 HCl,HF
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
定量限为检出限的3倍,需用空白加标实验验证实际定量能力
2. 至少10%平行样(相对偏差≤20%)
3. 加标回收率控制在80%-120%
4. 使用有证标准物质验证准确性
2. 含悬浮物样品需经0.45μm滤膜过滤
3. 内标加入(推荐Rh、Re、In混合内标)
4. 仪器调谐确保氧化物产率<3%
5. 采用标准曲线法(浓度范围覆盖0~100μg/L)
2. 注意多原子离子干扰(如40Ar35Cl对75As的干扰)
3. 高盐样品需稀释或使用碰撞反应池技术
ICP-OES法检出限0.01~5mg/kg,定量限按取样量换算
2. 平行样相对偏差≤10%
3. 加标回收率85%~110%
4. 使用标准物质(GBW100系列)监控
2. 赶酸至0.5mL以下
3. 定容后静置12小时以上
4. 内标校正(Ge/In/Bi混合溶液)
5. 标准曲线需包含0点、低、中、高浓度
2. 含钙量高样品注意基体匹配
3. 汞元素需单独优化仪器参数
AAS法检出限0.1~50μg/L(火焰法),0.001~0.1μg/L(石墨炉法)
2. 每批样品做10%平行双样
3. 每季度进行实验室间比对
2. ICP-MS分析前需用1%硝酸稀释10倍
3. 使用标准加入法校正高盐样品
4. 六价铬需用特定方法前处理
2. 砷、硒等元素建议使用氢化物发生法
3. 汞检测需用冷蒸气原子吸收法
以上信息仅供参考,请以相应标准的原文为准!