35种VOC混标/HJ 644-2013|VOC Mixture for HJ 644-2013 VOC Mixture for HJ 644-2013
-
HJ 642-2013 35种VOC混标|VOC Mixture for HJ 642-2013 VOC Mixture for HJ 642-2013
DRE-GA09000565ME | 1ml
-
35种VOC混标/HJ 644-2013|HJ 644-2013 VOC Mixture 532 HJ 644-2013 VOC Mixture 532
DRE-A50000532ME | 1ml
-
HJ 642-2013 8种VOC混标|VOC Mixture for HJ 642-2013 VOC Mixture for HJ 642-2013
DRE-GA09000552ME | 1ml
-
HJ 642-2013 60种VOC混标|VOC Mixture for HJ 642-2013 VOC Mixture for HJ 642-2013
DRE-GA09000551ME | 1ml
-
36种VOC混标/HJ 642-2013/HJ 643-2013|HJ 642-2013 VOC Mixture 136 HJ 642-2013 VOC Mixture 136
DRE-A50000136ME | 1ml
-
35种VOC混标/HJ 644-2013 VOC Mixture for HJ 644-2013 1000 µg/mL in Methanol
ZDR-GA09000562ME | 1ml
-
35种VOC混标/HJ 644-2013 HJ 644-2013 VOC Mixture 532 2000 µg/mL in Methanol
ZDR-A50000532ME | 1ml
产品类型:标准物质(RM)-溶液型
产品品牌:Dr.E
您正在浏览的产品:35种VOC混标/HJ 644-2013|VOC Mixture for HJ 644-2013
手机版:35种VOC混标/HJ 644-2013|VOC Mixture for HJ 644-2013
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
覆盖浓度范围:0.5 μg/m³~100 μg/m³
适用场景:环境空气背景值监测、污染源周边监测、室内空气质量评估
2. 前处理:热脱附仪(280-320℃)二次解吸,冷阱聚焦(-30℃)
3. 分析设备:气相色谱-质谱联用仪(GC-MS)
4. 色谱条件:DB-624等弱极性色谱柱,程序升温(初始40℃保持3min,以5℃/min升至120℃,再以10℃/min升至230℃)
5. 质谱条件:全扫描模式(SCAN),质量范围35-270 amu
2. 定量下限(LOQ):0.3 μg/m³~1.5 μg/m³
3. 相对响应因子(RRF)波动范围:±30%内有效
4. 标准曲线相关系数要求:R²≥0.995
2. 现场空白:每10个样品≥1个,全程同条件保存运输
3. 平行样:每批次≥10%样品量,相对偏差≤30%
4. 内标回收率:75%-125%(d4-甲苯等氘代化合物)
5. 标准曲线核查:每24小时用中间点校核,偏差≤20%
2. 样品采集:采样前记录温湿度气压,采样时间≤24小时
3. 样品保存:采样后两端密封,-20℃避光保存,14天内完成分析
4. 热脱附条件:一级脱附温度280℃,脱附时间5min;冷阱温度-30℃
5. 进样方式:分流比5:1~20:1,传输线温度150℃
6. 定性定量:目标物特征离子定性(≥3个离子),内标法定量
2. 水分干扰处理:在冷阱后安装Nafion®干燥管
3. 吸附管穿透验证:串联两支吸附管采样,后管浓度需<前管10%
4. 高浓度样品处理:采样体积减半或降低采样流量
5. 目标物丢失处理:含丙烯醛等极性组分需用Carbopack™复合吸附剂
6. 质谱调谐要求:定期进行PFK校准,满足DFTPP调谐标准
以上信息仅供参考,请以相应标准的原文为准!