壬烷中12种多溴联苯醚内标混标溶液(HJ 1270-2022-提取内标) 12 Stable Isotope Labeled PBDEs Mix Solution in Nonane
-
壬烷中3种多溴联苯醚内标混标溶液(HJ 1270-2022-进样内标) 3 Stable Isotope Labeled PBDEs Mix Solution in Nonane
1ST031244-VR1-N | 1mL
-
壬烷中3种多溴联苯醚内标混标溶液(HJ 1270-2022-进样内标) 3 Stable Isotope Labeled PBDEs Mix Solution in Nonane
1ST031244-VR2-N | 1mL
-
壬烷中13种多溴联苯醚内标混标溶液 13 Stable Isotope Labeled PBDEs Mix Solution in Nonane
1ST031031-VR1-N | 1mL
-
壬烷中2种多溴联苯醚内标混标溶液(HJ 1243-2022-进样内标),10μg/mL Stable Isotope Labeled 2 PBDEs Mix Solution in Nonane, 10μg/mL
1ST82048-10N | 1mL
-
甲苯中4种多溴联苯内标混标溶液(HJ 1243-2022-提取内标) 4 Stable Isotope Labeled PBBs Mix Solution in Toluene
1ST82047-VR1-T | 1mL
-
壬烷中8种多溴联苯醚内标混标溶液,5μg/mL 8 Stable Isotope Labeled PBDEs Mix Solution in Nonane,5μg/mL
1ST82818-5N | 1mL
-
壬烷中12种多氯联苯类内标混标溶液(GB 5009.205-2024附录B 表 B.5&B.7&HJ 1389-2024-提取内标),0.1μg/mL 12 Stable Isotope Labeled PCBs Mix Solution in Nonane, 0.1μg/mL
1ST82093-0.1N | 1mL
您正在浏览的产品:壬烷中12种多溴联苯醚内标混标溶液(-提取内标)
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
归口单位:生态环境部
2. 目标物:BDE-28、47、99、100、153、154、183、197、207、206、209等12种多溴联苯醚
3. 关键限定:需配合标准物质"壬烷中12种多溴联苯醚内标混标溶液"进行同位素稀释定量
固相萃取(SPE):HLB或C18柱富集,二氯甲烷洗脱
气相色谱-质谱联用(GC-MS)
色谱柱:DB-5HT (15m×0.25mm×0.1μm)
进样方式:不分流进样,程序升温(100℃→325℃)
检测模式:SIM选择离子监测
0.2~0.8 ng/L(不同目标物)
1. 标准曲线R²≥0.990
2. 内标回收率需在70%~130%范围
3. 十溴联苯醚(BDE-209)需特殊验证:降解率≤20%
每批次≤20个样品需做1个纯水空白,目标物响应值需<方法检出限
加标浓度:中浓度点(标准曲线中间值)
回收率范围:70%~130%(十溴联苯醚为50%~150%)
使用壬烷中12种多溴联苯醚内标混标溶液:
- 样品前处理前加入
- 相对响应因子RSD≤20%
- 全程监控提取效率
量取1000mL水样,加入50μL壬烷中12种多溴联苯醚内标混标溶液(HJ 1270-2022-提取内标)
HLB柱活化:5mL二氯甲烷→5mL甲醇→10mL纯水
上样流速:5mL/min,洗脱用8mL二氯甲烷
氮吹浓缩至0.5mL,转移至进样小瓶,加入回收率指示物
进样量1μL,采用时间窗口分段扫描,BDE-209单独设置监测参数
2. 溶剂效应规避:壬烷溶剂需与标准曲线基质一致
3. 污染防控:实验器皿禁用塑料材质,避免邻苯二甲酸酯干扰
4. 质谱调谐:每日需用
以上信息仅供参考,请以相应标准的原文为准!