14种VOC混标/HJ 620-2011 VOC mix for HJ 620-2011, various concentrations in Methanol
-
14种VOC混标/HJ 620-2011|VOC mix for HJ 620-2011, various concentrations VOC mix for HJ 620-2011, various concentrations
DRE-GA09000575ME | 1ml
-
HJ 644-2013 21种酚类混标 VOC mix 2 for HJ 644-2013, various concentrations in Methanol
ZDR-GA09000566ME | 1ml
-
HJ 644-2013 21种酚类混标(总效期:15个月)|VOC mix 2 for HJ 644-2013, various concentrations VOC mix 2 for HJ 644-2013, various concentrations
DRE-GA09000566ME | 1ml
-
11种VOC混标/HJ 686-2014|VOC mix2 for HJ 742-2015, VOC mix2 for HJ 742-2015,
DRE-GA09000571ME | 1ml
-
HJ 642-2013 60种VOC混标|VOC Mixture for HJ 642-2013 VOC Mixture for HJ 642-2013
DRE-GA09000551ME | 1ml
-
11种VOC混标/HJ 686-2014|HJ 686-2014 VOC Mixture 621 HJ 686-2014 VOC Mixture 621
DRE-A50000621ME | 1ml
-
37种VOC混标/HJ 741-2015/HJ 760-2015|VOC mix for HJ 760 -2015, VOC mix for HJ 760 -2015,
DRE-GA09000556ME | 1ml
中文名称:14种VOC混标/HJ 620-2011
英文名称:VOC mix for HJ 620-2011, various concentrations in Methanol
您正在浏览的产品:14种VOC混标/HJ 620-2011
手机版:14种VOC混标/HJ 620-2011
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
顶空进样-气相色谱法:水样在密封顶空瓶内经恒温平衡, 卤代烃从液相逸散至气相,取气相部分注入气相色谱仪分离检测。
检测器类型:推荐使用电子捕获检测器(ECD)或质谱检测器(MS)
方法检出限(MDL):0.1~0.5μg/L(不同组分有差异)
定量下限(LOQ):0.4~2.0μg/L(取检出限的4倍值)
典型组分示例:
三氯甲烷:MDL=0.2μg/L,LOQ=0.8μg/L
四氯化碳:MDL=0.1μg/L,LOQ=0.4μg/L
实验室空白:每批样品(≤20个)至少1个超纯水空白
基体加标:每批样品需做1组加标回收(加标浓度宜为待测物浓度的0.5~2倍)
平行样:至少10%样品做平行测定,相对偏差≤25%
标准曲线:至少5个浓度点,相关系数R²≥0.995
1. 样品采集:使用40mL棕色玻璃瓶零顶空采样,水样充满容器,立即密封冷藏(4℃)
2. 顶空平衡:取10mL水样至顶空瓶,60℃恒温平衡30min(振荡频率200r/min)
3. 色谱条件:DB-624毛细管柱(60m×0.32mm×1.8μm),程序升温(初始40℃保持5min,以5℃/min升至150℃)
4. 进样参数:顶空进样器传输线温度110℃,进样量1.0mL,分流比10:1
1. 干扰消除:余氯需用抗坏血酸消除(每40mL水样加25mg)
2. 样品保存:采样后14天内完成分析(添加盐酸至pH<2可延长至28天)
3. 防止污染:避免使用塑料制品接触样品,所有玻璃器皿需经400℃烘烤4h
4. 14种VOC混标:需包含三氯甲烷、四氯化碳、三氯乙烯、四氯乙烯等标准规定组分
以上信息仅供参考,请以相应标准的原文为准!