联苯菊酯、甲氰菊酯、氯氟氰菊酯等9种拟除虫菊酯类农药的测定。
当取样体积为1000 mL时,方法检出限为0.002 μg/L~0.005 μg/L。
2. 气相色谱分离(推荐DB-5MS色谱柱)
3. 质谱检测器(EI源)选择离子监测模式(SIM)分析
4. 保留时间和特征离子丰度比定性
5. 内标法或外标法定量
2. 加标回收率控制在70%~130%
3. 采用内标法时需添加氘代拟除虫菊酯内标
4. 标准曲线相关系数≥0.995
5. 每12小时进行中间浓度点校验(偏差≤20%)
2. 脱水:萃取液经无水硫酸钠柱脱水
3. 浓缩:氮吹浓缩至1.0 mL(控制温度≤40℃)
4. 仪器条件:进样口温度280℃,分流比10:1,程序升温(初始80℃,以20℃/min升至300℃)
5. 质谱参数:离子源温度230℃,传输线温度280℃
2. 玻璃器皿需硅烷化处理防止吸附
3. 浓缩过程需严格控制温度防止目标物损失
4. 当样品存在基质干扰时需采用基质匹配标准曲线
5. 含余氯水样需预先加入0.1 g抗坏血酸消除干扰