8种拟除虫菊酯类农药混标/HJ 753-2015|HJ 753-2015 Pyrethroid Pesticides Mixture 611 HJ 753-2015 Pyrethroid Pesticides Mixture 611
-
8种拟除虫菊酯类混标/HJ 753-2015|HJ 753-2015 Pyrethroid Pesticides Mixture 153 HJ 753-2015 Pyrethroid Pesticides Mixture 153
DRE-A50000153AC | 1ml
-
8种拟除虫菊酯类农药混标/HJ 753-2015 HJ 753-2015 Pyrethroid Pesticides Mixture 611 1000 µg/mL in Acetone
ZDR-A50000611AC | 1ml
-
8种拟除虫菊酯类混标/HJ 753-2015 HJ 753-2015 Pyrethroid Pesticides Mixture 153 100 µg/mL in Acetone
ZDR-A50000153AC | 1mL
-
9种除虫菊酯类农药混标/HJ 753-2015 Pyrethroid pesticides mix 9
MCPD190721 | 1mL
-
18种PCB混标/HJ 715-2014/HJ 743-2015/HJ 891-2017/HJ 922-2017|PCB Mixture for HJ 715-2014 / HJ 743-2015 PCB Mixture for HJ 715-2014 / HJ 743-2015
DRE-GA09000584TO | 1ml
-
18种PCB混标/HJ 715-2014/HJ 743-2015/HJ 891-2017/HJ 922-2017|PCB Mixture for HJ 743-2015 PCB Mixture for HJ 743-2015
DRE-GA09000583HE | 1ml
-
3种内标物混标/HJ 713-2014/HJ 714-2014/HJ 735-2015/HJ 736-2015 (氟苯、1-氯-2-溴丙烷、4-溴氟苯)|HJ 713, HJ 714-2014, HJ 735, HJ 736-2015 VOC Mixture 116 HJ 713, HJ 714-2014, HJ 735, HJ 736-2015 VOC Mixture 116
DRE-A50000116ME | 1ml
产品类型:标准物质(RM)-溶液型
产品品牌:Dr.E
您正在浏览的产品:8种拟除虫菊酯类农药混标/HJ 753-2015|HJ 753-2015 Pyrethroid Pesticides Mixture 611
手机版:8种拟除虫菊酯类农药混标/HJ 753-2015|HJ 753-2015 Pyrethroid Pesticides Mixture 611
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
发布机构:生态环境部
实施日期:2015年12月1日
• 目标化合物:联苯菊酯、甲氰菊酯、三氟氯氰菊酯、氯菊酯、氯氰菊酯、氰戊菊酯、溴氰菊酯
• 不适用范围:含油量>50 mg/L或悬浮物>500 mg/L的水样需特殊前处理
1. 水样前处理:C18或HLB固相萃取柱富集目标物
2. 洗脱溶剂:正己烷/丙酮(9:1)
3. 色谱条件:DB-5MS毛细管柱(30m×0.25mm×0.25μm)
4. 质谱模式:选择离子监测(SIM),电子轰击源(EI)
注:当取样量为1000mL时,其他菊酯类农药检出限范围1.5-3.0 ng/L
• 基质加标:每批样品至少1组,加标浓度LOQ的2-5倍
• 平行样:每批样品≥10%平行双样,相对偏差≤30%
• 标准曲线:线性范围LOQ~200μg/L,相关系数R2≥0.995
• 替代物回收率:使用氘代菊酯作为内标,回收率控制范围60-130%
2. 上样控制:流速保持3-5mL/min,避免柱床干涸
3. 氮吹浓缩:洗脱液氮吹至近干,正己烷定容至1.0mL
4. 仪器分析:进样量1.0μL,不分流进样,程序升温(初始80℃保持1min → 20℃/min升至290℃保持10min)
5. 定性定量:保留时间窗口±0.5min,特征离子丰度比±30%
• 含余氯水样需加入抗坏血酸(0.1g/100mL)
• 当氯菊酯出现双峰时,需峰面积加和定量
• 高盐度样品(电导>2000μS/cm)需稀释后萃取
• 方法验证需满足:空白无干扰,替代物回收率70-130%
以上信息仅供参考,请以相应标准的原文为准!