14种VOC混标/HJ 620-2011|VOC mix for HJ 620-2011, various concentrations VOC mix for HJ 620-2011, various concentrations
-
14种VOC混标/HJ 620-2011 VOC mix for HJ 620-2011, various concentrations in Methanol
ZDR-GA09000575ME | 1ml
-
HJ 644-2013 21种酚类混标(总效期:15个月)|VOC mix 2 for HJ 644-2013, various concentrations VOC mix 2 for HJ 644-2013, various concentrations
DRE-GA09000566ME | 1ml
-
11种VOC混标/HJ 686-2014|VOC mix2 for HJ 742-2015, VOC mix2 for HJ 742-2015,
DRE-GA09000571ME | 1ml
-
HJ 644-2013 21种酚类混标 VOC mix 2 for HJ 644-2013, various concentrations in Methanol
ZDR-GA09000566ME | 1ml
-
37种VOC混标/HJ 741-2015/HJ 760-2015|VOC mix for HJ 760 -2015, VOC mix for HJ 760 -2015,
DRE-GA09000556ME | 1ml
-
13种酚类混标/HJ 676-2013|Phenol Mixture for HJ 676-2013 various concentrations Phenol Mixture for HJ 676-2013 various concentrations
DRE-GA09000540ME | 1ml
-
HJ 642-2013 60种VOC混标|VOC Mixture for HJ 642-2013 VOC Mixture for HJ 642-2013
DRE-GA09000551ME | 1ml
产品类型:标准物质(RM)-溶液型
产品品牌:Dr.E
您正在浏览的产品:14种VOC混标/HJ 620-2011|VOC mix for HJ 620-2011, various concentrations
手机版:14种VOC混标/HJ 620-2011|VOC mix for HJ 620-2011, various concentrations
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
1. 顶空进样:水样在密封顶空瓶内经60℃恒温平衡
2. 色谱分离:DB-624毛细管色谱柱(60m×0.32mm×1.8μm)程序升温
3. 质谱检测:电子轰击源(EI)全扫描模式(SCAN)或选择离子扫描模式(SIM)
注:实际检出限需通过空白加标实验验证,当使用14种VOC混标时需匹配标准规定的浓度梯度。
1. 实验室空白:超纯水替代样品全程分析
2. 基质加标:实际样品添加标准物质(加标浓度≈定量下限5倍)
3. 平行样:≥10%样品做平行测定
4. 校准曲线:使用14种VOC混标建立,相关系数R²≥0.995
5. 连续校准:每20个样品插入中间浓度点校验,偏差≤15%
1. 取10mL水样注入顶空瓶,立即密封
2. 添加0.5g氯化钠(提高挥发性)
顶空条件:
平衡温度60℃±1℃,平衡时间30min
传输线温度110℃,加压压力0.08MPa
色谱条件:
初温40℃(保持4min),以5℃/min升至120℃,再以10℃/min升至220℃(保持2min)
质谱条件:
离子源温度230℃,扫描范围35~270amu
2. 干扰消除:余氯干扰需加入抗坏血酸(0.01g/mL)
3. 标准物质使用:14种VOC混标开封后需-18℃冷冻避光保存,使用前需平衡至室温
4. 系统维护:每200次进样后需更换进样口隔垫和衬管,定期切割色谱柱前端0.5m
以上信息仅供参考,请以相应标准的原文为准!