22种有机氯混标/HJ 921-2017(缺少艾氏剂)|HJ 921-2017 Organochlorine Pesticides Mixture 109 HJ 921-2017 Organochlorine Pesticides Mixture 109
-
22种有机氯混标/HJ 921-2017(缺少艾氏剂) HJ 921-2017 Organochlorine Pesticides Mixture 109 100 µg/mL in Toluene:n-Hexane
ZDR-A50000109TH | 1mL
-
23种有机氯混标/HJ 835-2017/HJ 900-2017/HJ 901-2017/HJ 912-2017|HJ 835-2017, HJ 835-2017, HJ 900-2017, HJ 901-2017, HJ 912-2017, HJ 921-2017 Organochlorine Pesticides Mixture 302 HJ 835-2017, HJ 835-2017, HJ 900-2017, HJ 901-2017, HJ 912-2017, HJ 921-2017 Organochlorine Pesticides Mixture 302
DRE-A50000302AH | 1ml
-
24种有机氯混标/HJ 835-2017/HJ 900-2017/HJ 901-2017/HJ 912-2017 (多一种三氯杀螨醇)|HJ 835-2017, HJ 900, HJ 901, HJ 912-2017 Organochlorine Pesticides Mixture105 HJ 835-2017, HJ 900, HJ 901, HJ 912-2017 Organochlorine Pesticides Mixture105
DRE-A50000105AH | 1ml
-
23种有机氯混标/HJ 835-2017/HJ 900-2017/HJ 901-2017/HJ 912-2017 HJ 835-2017, HJ 835-2017, HJ 900-2017, HJ 901-2017, HJ 912-2017, HJ 921-2017 Organochlorine Pesticides Mixture 302 1000 µg/mL in n-Hexane/Acetone
ZDR-A50000302AH | 1ml
-
有机氯农药混标 921|Organochlorine Pesticides Mixture 921 Organochlorine Pesticides Mixture 921
DRE-GA09000921AC | 1ml
-
2种替代物混标/HJ 699-2014/HJ 901-2017/HJ 903-2017/HJ 904-2017(2,4,5,6-四氯间二甲苯、十氯联苯)|HJ 699-2014, HJ 904-2017, HJ 901-2017, HJ 903-2017, HJ 743-2015 Pesticide Surrogate Mixture 489 HJ 699-2014, HJ 904-2017, HJ 901-2017, HJ 903-2017, HJ 743-2015 Pesticide Surrogate Mixture 489
DRE-A50000489TH | 1ml
-
15种氨基甲酸酯类混标 HJ 961-2018/HJ 1026-2019|HJ 961-2018, HJ 1026-2019 Pesticides Mixture 497 HJ 961-2018, HJ 1026-2019 Pesticides Mixture 497
DRE-A50000497ME | 1ml
产品类型:标准物质(RM)-溶液型
产品品牌:Dr.E
您正在浏览的产品:22种有机氯混标/HJ 921-2017(缺少艾氏剂)|HJ 921-2017 Organochlorine Pesticides Mixture 109
手机版:22种有机氯混标/HJ 921-2017(缺少艾氏剂)|HJ 921-2017 Organochlorine Pesticides Mixture 109
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
1. 色谱柱:DB-5MS (30m×0.25mm×0.25μm)或等效柱
2. 进样方式:不分流进样
3. 离子源:电子轰击源(EI)
4. 扫描模式:选择离子监测(SIM)
2. 加标回收率控制在70%~130%
3. 使用有证标准物质(CRM)验证,频次≥5%
4. 保留时间变化≤±0.1min
1. 索氏提取:10g样品+100mL丙酮/正己烷(1:1)混合溶剂
2. 净化:硅胶柱层析,用15mL正己烷淋洗后,30mL二氯甲烷/正己烷(3:7)洗脱
3. 浓缩:氮吹仪浓缩至0.5mL,转移至进样小瓶
4. 上机分析:GC-MS进样量1μL
2. 艾氏剂缺失处理:需单独配制艾氏剂标准溶液补充
3. 干扰消除:样品需经铜粉除硫处理
4. 色谱峰识别:要求目标物与干扰物分离度R≥1.0
以上信息仅供参考,请以相应标准的原文为准!