64种SVOCs混标/HJ 834-2017/HJ 951-2018|CLP SVOC Calibration Mixture 512 CLP SVOC Calibration Mixture 512
-
64种SVOCs混标/HJ 834-2017/HJ 951-2018 (总效期:18个月)|HJ 834-2017, HJ 951-2018 VOC & SVOCs Mixture 155 HJ 834-2017, HJ 951-2018 VOC & SVOCs Mixture 155
DRE-A50000155DI | 1ml
-
6种替代物混标/HJ 834-2017/HJ 951-2018|HJ 834-2017, HJ 951-2018 VOC & SVOCs Substitutes Mixture 156 HJ 834-2017, HJ 951-2018 VOC & SVOCs Substitutes Mixture 156
DRE-A50000156AI | 1ml
-
64种SVOCs混标/HJ 834-2017/HJ 951-2018 SVOCs Mix 64
MCEM591010-DI | 1mL
-
64种SVOCs混标/HJ 834-2017/HJ 951-2018 HJ 834-2017, HJ 951-2018 VOC & SVOCs Mixture 155 1000 µg/mL in Dichloromethane
ZDR-A50000155DI | 1mL
-
4种SVOC混标/HJ 801-2016|HJ 801-2016 SVOC Mixture 536 HJ 801-2016 SVOC Mixture 536
DRE-A50000536WA | 1ml
-
6种内标物混标/HJ 834-2017/HJ 951-2018|HJ 834-2017, HJ 951-2018 VOC & SVOCs internal standards HJ 834-2017, HJ 951-2018 VOC & SVOCs internal standards
DRE-A50000174AI | 1ml
-
3种内标物混标/HJ 835-2017|SVOC Mixture 623 SVOC Mixture 623
DRE-A50000623DI | 1ml
产品类型:标准物质(RM)-溶液型
产品品牌:Dr.E
您正在浏览的产品:64种SVOCs混标/HJ 834-2017/HJ 951-2018|CLP SVOC Calibration Mixture 512
手机版:64种SVOCs混标/HJ 834-2017/HJ 951-2018|CLP SVOC Calibration Mixture 512
本公司销售的所有产品仅供实验科研使用,不用于人体及临床诊断。
1. 色谱柱:5%苯基甲基聚硅氧烷毛细管柱(30m×0.25mm×0.25μm)
2. 进样方式:不分流进样
3. 离子源:电子轰击源(EI)
2. 定量下限:0.08~2.4 mg/kg
3. 需通过实际样品验证
2. 每20个样品做1个平行样(相对偏差≤30%)
3. 每批样品做1个基体加标样(回收率控制范围50%~130%)
4. 每半年验证校准曲线
2. 萃取溶剂:丙酮-正己烷(1:1)混合溶液
3. 净化处理:硅胶柱或弗罗里硅土柱净化
4. 浓缩定容:氮吹浓缩至1.0mL
5. 仪器分析:GC-MS选择离子扫描模式(SIM)
2. 酞酸酯类需防止实验室塑料器皿污染
3. 含硫样品需铜粉除硫处理
4. 高有机质样品需增加净化步骤
1. 色谱柱:DB-5MS(30m×0.25mm×0.25μm)
2. 程序升温:初始40℃(保持4min)→285℃(10℃/min)
3. 检测模式:全扫描(SCAN)或选择离子监测(SIM)
2. 定量下限:2.0~40 μg/kg
3. 实际样品需考虑基体效应
2. 每10个样品做1个平行样(RSD≤25%)
3. 每批样品做基体加标回收(70%~130%)
4. 定期进行实验室间比对
2. 提取方式:加压流体萃取(PFE)或超声波萃取
3. 萃取溶剂:二氯甲烷-丙酮(1:1)
4. 浓缩条件:旋转蒸发仪浓缩(水浴≤40℃)
5. 净化方法:凝胶渗透色谱(GPC)联合硅胶柱
2. 含油样品需增加氧化铝净化
3. 不稳定化合物需衍生化处理
4. 检测报告需注明所用定量方法
2. 每日分析前需进行中间浓度点核查(偏差≤20%)
3. 内标法定量(添加氘代内标物)
2. 保留时间窗口:±0.05min
3. 离子丰度比:±20%允许范围
以上信息仅供参考,请以相应标准的原文为准!